Klarity Medical Products LLC recalls Klarity Vacuum Bags, Item Nos. R7504-30C, R7504-35BC,R7504-35C, R7644-32.5BC, RC7221-25BC, RC7504-17.5C - Product Usage…
- Recall date
- July 31, 2020
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2881-2020
- FDA classification
- Class II
- Brand / firm
- Klarity Medical Products LLC
- Sold / distributed
- Worldwide distribution - US Nationwide distribution including in the states of CA, FL, MA, NY, PA and the country of UK.
Why it was recalled
Vacuum bags potentially may not hold their vacuum adequately for needed patient treatment times.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Klarity Vacuum Bags, Item Nos. R7504-30C, R7504-35BC,R7504-35C, R7644-32.5BC, RC7221-25BC, RC7504-17.5C - Product Usage: To be used by trained medical professionals for the stable support and positioning of patients undergoing external beam radiation therapy treatment in a clinic or hospital setting.
Get recall alerts
Free email alert whenever Klarity Medical Products LLC has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Klarity Medical Products LLC