Medical device recalls Moderate risk

Konica Minolta Medical Imaging USA, Inc. recalls Sonimage HS1 Ultrasound Kit AC adapter, Product Number: 1. A9RP, UDI: 04560141945463; 2. AAEU, UDI: 04560141946385; 3.…

Recall date
December 4, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1214-2018
FDA classification
Class II
Brand / firm
Konica Minolta Medical Imaging USA, Inc.
Sold / distributed
Distributed in the following U.S. states: AL, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, WA, WI and the District of Columbia. Distributed in Canada and Brazil.

Why it was recalled

There is a risk of circuit board breakage within the main housing of the AC Adapter unit. Smoke may be emitted from the AC Adapter as a direct result.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Sonimage HS1 Ultrasound Kit AC adapter, Product Number: 1. A9RP, UDI: 04560141945463; 2. AAEU, UDI: 04560141946385; 3. A9RR, UDI: 04560141945470