Medical device recalls Moderate risk

Kreatech Biotechnology Bv recalls Kreatech PapV-16, Flu labeled ISH probe, to detect presence of HPV-16 DNA inserts in human tissue, Model Number 40V016V…

Recall date
November 1, 2024
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0522-2025
FDA classification
Class II
Brand / firm
Kreatech Biotechnology Bv
Sold / distributed
US States: FL, NY

Why it was recalled

The DNA fragments for this specific lot number are too large, resulting in potential background around the cells due to suboptimal hybridization. An absence of signal could also be observed as a result of the suboptimal hybridization of the larger DNA fragments. The user may interpret potential background as a false positive result and an absence of signal as a false negative result, if proper positive and negative controls are not included.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Kreatech PapV-16, Flu labeled ISH probe, to detect presence of HPV-16 DNA inserts in human tissue, Model Number 40V016V495