Laboratoires Bodycad, Inc. recalls Torque Limiting Adaptor, AO Connect, 4 N-m, part #027-203, reusable, a component of the Bodycad Fine Osteotomy Implant…
- Recall date
- July 2, 2021
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0244-2022
- FDA classification
- Class II
- Brand / firm
- Laboratoires Bodycad, Inc.
- Sold / distributed
- Distribution was made to CO, MI, and NY. Foreign distribution was made to Canada and the United Kingdom. There was no military/government distribution.
Why it was recalled
There is a risk of breakage of the torque limiting adaptor if used counterclockwise and it must not be used with a power tool due to no test data validating this use.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Torque Limiting Adaptor, AO Connect, 4 N-m, part #027-203, reusable, a component of the Bodycad Fine Osteotomy Implant and Disposable Instrumentation Tray, part #025-004.