Laerdal Medical Corporation recalls Laerdal Compact Suction Unit 4 RTCA version (LCSU 4 RTCA) 1) LCSU 4, 800 ml, RTCA; Catalog Number: 880052; 2) LCSU 4, 3…
- Recall date
- January 13, 2026
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1341-2026
- FDA classification
- Class II
- Brand / firm
- Laerdal Medical Corporation
- Sold / distributed
- Worldwide distribution - US Nationwide and the countries of Denmark, Sweden, Norway, Canada, Finland, Australia, Spain, France, United Kingdom, Italy, Korea, Germany, Netherlands, Switzerland, New Zealand, Singapore, Malaysia, Austria.
Why it was recalled
units manufactured in a limited time period may emit electromagnetic noise beyond the acceptable limits for RTCA application specified in RTCA DO-160G Chapter 21, Equipment Category M.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Laerdal Compact Suction Unit 4 RTCA version (LCSU 4 RTCA) 1) LCSU 4, 800 ml, RTCA; Catalog Number: 880052; 2) LCSU 4, 300 ml, RTCA; Catalog Number: 880062; 3) LCSU 4, RTCA; Catalog Number: 881152;
Get recall alerts
Free email alert whenever Laerdal Medical Corporation has a new recall — straight from official government data. Unsubscribe anytime.