Landauer recalls NanoDots, Model Numbers: a) BC30023, description: QC Set For Microstar Dots; b) BC30083, description: nanoDot QC Set -…
- Recall date
- July 12, 2023
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2573-2023
- FDA classification
- Class II
- Brand / firm
- Landauer
- Sold / distributed
- Worldwide
Why it was recalled
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
NanoDots, Model Numbers: a) BC30023, description: QC Set For Microstar Dots; b) BC30083, description: nanoDot QC Set - 80kVp - Sale; c) BC30084, description: nanoDot QC Set - 80kVp - No Charge; d) BC30088, description: nanoDot Calibration Set - 80kVp - No Charge; e) BC30095, description: Dot, D1DNN, in carrier sealed in bag; f) BC30110, description: Calibration set, nanoDot; g) BC30141, description: nanoDot QC Set - Cs137 - Sale; h) BC30142, description: nanoDot Calibration set - Cs137 - Sale; radiation monitoring dosimeter used with the microSTAR readers