Medical device recalls Moderate risk

LeMaitre Vascular, Inc. recalls AnastoClip GC Closure System, REF# 4007-08 a closure system designed to penetrate the outer layers of tissue during the…

Recall date
June 19, 2018
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-3042-2018
FDA classification
Class II
Brand / firm
LeMaitre Vascular, Inc.
Sold / distributed
Distributed to accounts in FL, NV, OK, SC, and TX.

Why it was recalled

The wrong instructions for use (IFU) were supplied with the AnastoClip GC closure system. The IFU provided lists some indications which are not cleared by the FDA for devices distributed in the United States: 1) closure of arteriotomies and veinotomies, 2) attachment of synthetic vascular prostheses, and 3) approximation of dural tissue.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

AnastoClip GC Closure System, REF# 4007-08 a closure system designed to penetrate the outer layers of tissue during the forming of an anastomosis or closing of a vessel with possible intraluminal penetration.