Skintact Electrode for Defibrillation recalled over sterility concerns
- Recall date
- August 8, 2017
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Official notice title
- Leonhard Lang Medizintechnik GmbH recalls Skintact Electrode for Defibrillation, Pacing, Monitoring, Cardioversion. Product Usage: Multifunction electrode for ex…
- Recall number
- Z-3109-2017
- FDA classification
- Class I
- Brand / firm
- Leonhard Lang Medizintechnik GmbH
- Sold / distributed
- Worldwide distribution. US distribution in the states of : FL & TX. Austria, China, France, Germany, Greece, Hungary, India, Italy, Panama, Philippines, Poland, Saudi Arabia, Serbia, South Africa, South Korea, Spain, United Arab Emirates, United Kingdom.
Why it was recalled
Certain lots of the defibrillation electrodes model DF59N and DF59NC may have a delay in connecting or not connect at all with the defibrillator ZOLL AED PLUS. This may cause a situation in which a patient, who is in a life threatening condition and requires a defibrillation shock, cannot be treated in good time.
Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.
What was recalled
Skintact Electrode for Defibrillation, Pacing, Monitoring, Cardioversion. Product Usage: Multifunction electrode for external defibrillation, pacing, cardioversion, and monitoring. The device is non-sterile and for single use only. Used with automated external defibrillators
Get recall alerts
Free email alert whenever Leonhard Lang Medizintechnik GmbH has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Leonhard Lang Medizintechnik GmbH