LEONI CIA CABLE SYSTEMS recalls ORION System with software OSS v6.0 and OSS v7.0- An electro-mechanical robotic arm for patient positioning in Radiatio…
- Recall date
- October 6, 2023
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0398-2024
- FDA classification
- Class II
- Brand / firm
- LEONI CIA CABLE SYSTEMS
- Sold / distributed
- US Nationwide distribution in the states of MA, TN.
Why it was recalled
Users of the patient positioning system ORION System can be faced with a sudden short-term acceleration or with an unintended short-distance movement when the movements are again allowed after an unforeseen interruption signal of motion may result in a collision of the patient with another stationary component of the treatment room.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
ORION System with software OSS v6.0 and OSS v7.0- An electro-mechanical robotic arm for patient positioning in Radiation Therapy, Radiology, or Radiology image-guided treatment environment. Part Number: 1SOPT00003
Get recall alerts
Free email alert whenever LEONI CIA CABLE SYSTEMS has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: LEONI CIA CABLE SYSTEMS