Medical device recalls Moderate risk

Life Technologies Holdings Pte Ltd recalls Torrent Suite Dx Software Version (IUO) 5.4 Catalog Number: A31774 OEM Customer Software: Sentosa SQ Suite Software v5.…

Recall date
September 26, 2019
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0186-2020
FDA classification
Class II
Brand / firm
Life Technologies Holdings Pte Ltd
Sold / distributed
states : AZ, CA, CO, CT, FL, GA, IN, MA, MD, NC, NJ, NM, NY, OR, TX, VA, WA OUS: South Korea, ITALY, China, Japan, Singapore, Germany, TAIWAN, Poland

Why it was recalled

Torrent SuiteTM Dx analysis software, a component of the Ion PGM Dx Instrument System, incorrectly displays a positive result when a different analytical or de novo variant is detected at the same locus as a specified clinical variant

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Torrent Suite Dx Software Version (IUO) 5.4 Catalog Number: A31774 OEM Customer Software: Sentosa SQ Suite Software v5.6.15

Get recall alerts

Free email alert whenever Life Technologies Holdings Pte Ltd has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Life Technologies Holdings Pte Ltd