Life Technologies Holdings Pte Ltd recalls Torrent Suite Dx Software Version 5.6.4 Catalog Number:A33178 OEM Customer software: Sentosa SQ Suite Software v5.6.20…

Recall date
September 26, 2019
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0187-2020
FDA classification
Class II
Brand / firm
Life Technologies Holdings Pte Ltd
Sold / distributed
states : AZ, CA, CO, CT, FL, GA, IN, MA, MD, NC, NJ, NM, NY, OR, TX, VA, WA OUS: South Korea, ITALY, China, Japan, Singapore, Germany, TAIWAN, Poland

Why it was recalled

Torrent SuiteTM Dx analysis software, a component of the Ion PGM Dx Instrument System, incorrectly displays a positive result when a different analytical or de novo variant is detected at the same locus as a specified clinical variant

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Torrent Suite Dx Software Version 5.6.4 Catalog Number:A33178 OEM Customer software: Sentosa SQ Suite Software v5.6.20 UDI: (01)10190302005654 (11)000000 (10)5.6.4 (240)A33178