Medical device recalls Moderate risk

LivaNova Deutschland GmbH recalls Control Panel REF - 60-02-50: LivaNova SCP Pump Control Panel S5/C5. REF 60-02-50, IPX2, RxOnly, UDI: 04033817900993

Recall date
August 26, 2019
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2439-2020
FDA classification
Class II
Brand / firm
LivaNova Deutschland GmbH
Sold / distributed
US: FL, GA, IN, KY, MI, MO, NY, OH, PA, SC, TN, TX, VA OUS: Canada - Uzbekistan - Turkey - Tunisia - Taiwan - South Africa - Serbia - Saudi Arabia - Russia - Poland - Mexico - Korea(Rep. of) - India - Hong Kong - Colombia - China - Brazil - Bahrain - Argentina - Algeria - Italy - Germany

Why it was recalled

The SCP control Panel rotary knob used to control rotary speed of the pump in the SCP system used as a centrifugal pump during cardiopulmonary bypass may be difficult or impossible to rotate as a result of premature wear of the angle encoder. The loss of function of the angle encoder during bypass may lead to ischemia, stroke, organ damage, arterial tissue injury or reperfusion syndrome.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Control Panel REF - 60-02-50: LivaNova SCP Pump Control Panel S5/C5. REF 60-02-50, IPX2, RxOnly, UDI: 04033817900993