Medical device recalls Moderate risk

Lumitex Inc recalls Giraffe Blue Spot PT LiteTM Phototherapy System, Model Numbers M1224917, M1231533, M1225025, M1224961, M1224958, M12249…

Recall date
December 11, 2019
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0923-2020
FDA classification
Class II
Brand / firm
Lumitex Inc
Sold / distributed
The products were distributed to the following US states: WI.

Why it was recalled

There is a potential for some systems built between 2014 and Jan 2018 to have a light output falling below the recommended minimum output. This problem has been associated with exposure of the light pipe to elevated temperatures, which is contra-indicated in the Operation, Maintenance, and Service Manual.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Giraffe Blue Spot PT LiteTM Phototherapy System, Model Numbers M1224917, M1231533, M1225025, M1224961, M1224958, M1224957, M1224959, M1224963, M1225773, M1225066, M1224960, M1225059, M1225023, M1225060, M1225063, M1225064, M1225024, M1231622, M1236303, M1238489, M1238490, M1238492, M1238493, M1238495, M1238499, M1238503, M1238505, M1238507, M1238508, 2063860-001, 2063862-001, 2063870-001, 2063873-001, 2063874-001, 2071393-001, 2080916-001, 2089424-001