Maquet Cardiopulmonary Gmbh recalls Bubble Sensor for 3/8" x 3/32" tubing; Model Number: 701055720;
- Recall date
- January 9, 2026
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1300-2026
- FDA classification
- Class II
- Brand / firm
- Maquet Cardiopulmonary Gmbh
- Sold / distributed
- Worldwide Distribution: US (Nationwide) and OUS (International) to countries of: Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, E…
Why it was recalled
Internal investigations have identified an issue with the durability of the connecting cable near the connection to the Venous Bubble Sensor. Excessive bending of the connecting cable can lead to full damage of the Venous Bubble Sensor or loose cable contact, which may trigger the errors Ven. bubble sensor defective or Ven. bubble sensor disconnected on the connected medical device. These errors can occur temporarily when the connecting cable is moved or permanently if the connection is fully compromised.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Bubble Sensor for 3/8" x 3/32" tubing; Model Number: 701055720;
Get recall alerts
Free email alert whenever Maquet Cardiopulmonary Gmbh has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Maquet Cardiopulmonary Gmbh