Medical device recalls Moderate risk

Maquet Cardiovascular, LLC recalls Maquet Getinge: (1) Material: 701053941R01 BEQ-TOP 9005 3/8 Custom Pk w/RF32 & HMOD (2) Material: 701053941R02- BEQ-TOP…

Recall date
April 18, 2019
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1646-2019
FDA classification
Class II
Brand / firm
Maquet Cardiovascular, LLC
Sold / distributed
Nationwide

Why it was recalled

Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Maquet Getinge: (1) Material: 701053941R01 BEQ-TOP 9005 3/8 Custom Pk w/RF32 & HMOD (2) Material: 701053941R02- BEQ-TOP 9005 3/8 Custom Pk w RF32 & HMOD Material:

Get recall alerts

Free email alert whenever Maquet Cardiovascular, LLC has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Maquet Cardiovascular, LLC