Maquet Cardiovascular, LLC recalls Maquet Getinge-BEQ-TOP 32200 ECC PACK - Material: 701053319
- Recall date
- April 18, 2019
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1524-2019
- FDA classification
- Class II
- Brand / firm
- Maquet Cardiovascular, LLC
- Sold / distributed
- Nationwide
Why it was recalled
Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Maquet Getinge-BEQ-TOP 32200 ECC PACK - Material: 701053319
Get recall alerts
Free email alert whenever Maquet Cardiovascular, LLC has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Maquet Cardiovascular, LLC