Maquet Cardiovascular Us Sales product recalled over fire hazard
- Recall date
- November 13, 2015
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Official notice title
- Maquet Cardiovascular Us Sales, Llc recalls Maquet HCU 30 Coolant Type R134A Filling weight 250 g 120 V 50-60 Hz 1920 W Product Usage: The Heater Cooler Unit (HCU)…
- Recall number
- Z-0589-2016
- FDA classification
- Class II
- Brand / firm
- Maquet Cardiovascular Us Sales, Llc
- Sold / distributed
- United States Nationwide Distribution
Why it was recalled
Failure of the HEATER-COOLER UNIT (HCU 30) Type 2 power supply board. A drop of more than 10% voltage can cause overheating and subsequent failing of the R56 thermistor of the power supply board. Board failure typically occurs when the compressor is switched on (prior to patient use); however, it can occur during patient use. This can shut down the compressor or the entire HCU 30 Type 2 unit.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Maquet HCU 30 Coolant Type R134A Filling weight 250 g 120 V 50-60 Hz 1920 W Product Usage: The Heater Cooler Unit (HCU) 30 is intended to circulate water through heat exchange circuits to warm or cool a patient during short duration cardiopulmonary bypass procedures lasting 6 hours or less.
Get recall alerts
Free email alert whenever Maquet Cardiovascular Us Sales, Llc has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Maquet Cardiovascular Us Sales, Llc