Medical device recalls Moderate risk

Maquet Cardiovascular Us Sales product recalled over fire hazard

Recall date
November 13, 2015
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Official notice title
Maquet Cardiovascular Us Sales, Llc recalls Maquet HCU 30 Coolant Type R134A Filling weight 250 g 120 V 50-60 Hz 1920 W Product Usage: The Heater Cooler Unit (HCU)…
Recall number
Z-0589-2016
FDA classification
Class II
Brand / firm
Maquet Cardiovascular Us Sales, Llc
Sold / distributed
United States Nationwide Distribution

Why it was recalled

Failure of the HEATER-COOLER UNIT (HCU 30) Type 2 power supply board. A drop of more than 10% voltage can cause overheating and subsequent failing of the R56 thermistor of the power supply board. Board failure typically occurs when the compressor is switched on (prior to patient use); however, it can occur during patient use. This can shut down the compressor or the entire HCU 30 Type 2 unit.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Maquet HCU 30 Coolant Type R134A Filling weight 250 g 120 V 50-60 Hz 1920 W Product Usage: The Heater Cooler Unit (HCU) 30 is intended to circulate water through heat exchange circuits to warm or cool a patient during short duration cardiopulmonary bypass procedures lasting 6 hours or less.

Get recall alerts

Free email alert whenever Maquet Cardiovascular Us Sales, Llc has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Maquet Cardiovascular Us Sales, Llc