Medical device recalls High risk

Maquet Cardiovascular Us Sales, Llc recalls QUADROX-i Neonatal without integrated arterial filter and VHK 11000 (VKMO 10000-USA#QUADROX-i Neo.+Reservoir), Part Num…

Recall date
October 25, 2018
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0436-2019
FDA classification
Class I
Brand / firm
Maquet Cardiovascular Us Sales, Llc
Sold / distributed
Worldwide Distribution: US (nationwide) to states of: AZ, CA, FL, IA, LA, NC, and TX and countries of: Canada, Argentina, Brazil, Chile, Colombia, Mexico, Uruguay, United Arab Emirates, Belgium, Switzerland, Germany, Egypt, Spain, France, United Kingdom, Ireland, Italy, Lebanon, Netherlands, Pakist…

Why it was recalled

There is a potential for blood leakage at the blood inlet connector when using the reducing screw adapter 1/4 x 3/16 that may result in clinically relevant loss of prime before patient connection or blood loss during patient connection.

Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.

What was recalled

QUADROX-i Neonatal without integrated arterial filter and VHK 11000 (VKMO 10000-USA#QUADROX-i Neo.+Reservoir), Part Number 701055120 The membrane oxygenator QUADROXi Neonatal is intended for use in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery.

Get recall alerts

Free email alert whenever Maquet Cardiovascular Us Sales, Llc has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Maquet Cardiovascular Us Sales, Llc