QUADROX-i Pediatric recalled over sterility concerns
- Recall date
- May 18, 2023
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Official notice title
- Maquet Medical Systems USA recalls QUADROX-i Pediatric, Oxygenator used during cardiopulmonary bypass Model Item No. HMO 30000-USA 701070384 HMO 31000-USA…
- Recall number
- Z-1655-2023
- FDA classification
- Class II
- Brand / firm
- Maquet Medical Systems USA
- Sold / distributed
- Nationwide domestic distribution
Why it was recalled
Product packaging may exhibit small pinholes which may not be visible to the unaided eye, compromising the sterile barrier of the packaging. Exposure to non-sterile device may result in inflammation, infection, sepsis, and ischemia.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
QUADROX-i Pediatric, Oxygenator used during cardiopulmonary bypass Model Item No. HMO 30000-USA 701070384 HMO 31000-USA 701070388
Get recall alerts
Free email alert whenever Maquet Medical Systems USA has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Maquet Medical Systems USA