Medical device recalls Moderate risk

Maquet Medical Systems USA recalls Venous probe connection cable (REF 701069333 and 701048804) used with the Cardiohelp-i System, part numbers 701072780 a…

Recall date
November 13, 2023
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0605-2024
FDA classification
Class II
Brand / firm
Maquet Medical Systems USA
Sold / distributed
Domestic distribution Nationwide. Foreign distribution to Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Canada, Cayman Islands, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Estonia, Finland, France, French Polynesia, Germany, Greece, Hong Kong, Hungary, Iceland, India, I…

Why it was recalled

Firm received complaints of breaks in the insulation, shielding, or wires of the venous probe connection cable. Fracture of the venous probe cable may influence the management of extracorporeal support. Patient may be exposed to inappropriate blood flow, blood temperature, or hematocrit/hemoglobin levels as a result of the CARDIOHELP-i not receiving information from the venous probe.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Venous probe connection cable (REF 701069333 and 701048804) used with the Cardiohelp-i System, part numbers 701072780 and 701041802

Get recall alerts

Free email alert whenever Maquet Medical Systems USA has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Maquet Medical Systems USA