Medical device recalls Moderate risk

Med Tec Inc recalls Protura Robotic Couch System Software, versions 1.6.1 (Part Number MT6XSM1.6.1), 1.7.0 (Part Number MT6XSM1.7.0), and 1…

Recall date
March 31, 2016
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1992-2016
FDA classification
Class II
Brand / firm
Med Tec Inc
Sold / distributed
Worldwide Distribution - US Distribution including the countries of Switzerland, Germany..

Why it was recalled

When using the recalled product in split shift mode, the non-sub-millimeter positioning has been found to be inconsistent with the setting parameters.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Protura Robotic Couch System Software, versions 1.6.1 (Part Number MT6XSM1.6.1), 1.7.0 (Part Number MT6XSM1.7.0), and 1.7.1 (Part Number MT6XSM1.7.1), CIVCO, Orange City, IA 51041

Get recall alerts

Free email alert whenever Med Tec Inc has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Med Tec Inc