Medical device recalls Moderate risk

Medica Corporation recalls Wash Kit (REF 10680) Lots 16279, 17045 and 17215 Medica corporation is marketing a wash solution Chemistry Wash1 , REF…

Recall date
November 21, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1149-2018
FDA classification
Class II
Brand / firm
Medica Corporation
Sold / distributed
Worldwide and US Distribution.

Why it was recalled

The firm has initiated a field recall of the Wash1 Kit REF 10680 due to reported inconsistent Lipase values when analyzing patient samples. The firm's internal investigation identified the underlying cause to be an ineffective wash solution resulting in an interference carryover from the triglyceride (TRIG) assay if they are on the same worklist.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Wash Kit (REF 10680) Lots 16279, 17045 and 17215 Medica corporation is marketing a wash solution Chemistry Wash1 , REF # 10680 for use with Medica s EasyRA clinical chemistry analyzer. The solution is used prior to a lipase (LIP) assay if an interferant assay (TRIG)is part of the work list. The major interferant for Lipase is the Triglycerides (TRIG) assay since one of the reagents contains lipase as an active ingredient.