Medicalplastic S.R.L. recalls Ilizarov Wire Tensioner, Model 71070341, UDI (10)08033201842072(10)21HP02550(11)210831(21E434
- Recall date
- June 17, 2021
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0245-2022
- FDA classification
- Class II
- Brand / firm
- Medicalplastic S.R.L.
- Sold / distributed
- Memphis, TN
Why it was recalled
Medicalplastic s.r.l. is conducting a voluntary field action to remove a group of Ilizarov Wire Tensioners due to a manufacturing error. An internal washer of the Ilizarov Wire Tensioner has been made in a wrong plastic material that could wear and break during the sterilization process in autoclave. Consequently this could cause malfunctioning of the tensioner.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Ilizarov Wire Tensioner, Model 71070341, UDI (10)08033201842072(10)21HP02550(11)210831(21E434
Get recall alerts
Free email alert whenever Medicalplastic S.R.L. has a new recall — straight from official government data. Unsubscribe anytime.