Medical device recalls Moderate risk

Medistim Asa recalls Brand Name: Medistim MiraQ System Product Name: Medistim MiraQ System Model/Catalog Number: MQC04001, MQC14001, MQU1401…

Recall date
September 26, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0410-2026
FDA classification
Class II
Brand / firm
Medistim Asa
Sold / distributed
Domestic distribution to CO, MI, MN, TN, TX. International distribution to Australia, Austria, Germany, Finland, India, Japan, Russia, China, Saudi Arabia.

Why it was recalled

Printer Support Module (PSM) may be subject to insufficient soldering, which may cause loosening of the wires and system to shut down. Affected devices need to have their Printer Support Modules replaced.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Brand Name: Medistim MiraQ System Product Name: Medistim MiraQ System Model/Catalog Number: MQC04001, MQC14001, MQU14011, MQV12011 Software Version: Not applicable Product Description: Ultrasound device for intraoperative use Component: No

Get recall alerts

Free email alert whenever Medistim Asa has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Medistim Asa