Medical device recalls Moderate risk

Medivators, Inc. recalls Medivators EndoGator REF 100651. Kit for Olympus 140/160/180/190 Series GI Endoscopes for CO2 Connection with Endogator…

Recall date
October 9, 2019
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1777-2020
FDA classification
Class II
Brand / firm
Medivators, Inc.
Sold / distributed
Worldwide distribution - US Nationwide distribution and the countries of Australia, England.

Why it was recalled

Medivators is recalling Endo SmartCap Tubing due to a manufacturing anomaly that may present during procedures as inadequate insufflation, presence of air bubbles inside the endo smartcap tubing, insufficient water to rinse the endoscope lens, or water leaking from the air/water valve.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Medivators EndoGator REF 100651. Kit for Olympus 140/160/180/190 Series GI Endoscopes for CO2 Connection with Endogator EGP-100, Olympus OFP, Olympus AFU-100 & Erbe EIP 2 Irrigation Pump Units. Contents 10. Sterile EO. of 100145CO2EXT Endo SmartCap Irrigation Tubing for Olympus 140/160/180/190 Series GI Endoscopes with Extended CO2 Input - Product Usage: is intended to be used with an air or CO2 source and/or a pump along with a sterile water source to supply air or CO2 and sterile water to an endoscope during endoscopic procedures.