Medline Industries Inc recalls Centurion Convenience Kits, containing BD Alaris Pump Infusion Sets. BD code 11522558 (Centurion code 110931) contained…
- Recall date
- May 10, 2019
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2067-2019
- FDA classification
- Class I
- Brand / firm
- Medline Industries Inc
- Sold / distributed
- One distributor located in Illinois.
Why it was recalled
This recall is a sub recall to BD s recent Class I recall for the Alaris Pump Infusion Set due to an issue with the consistency of the inner lumen thickness of the silicone tubing segment which directly impacts amount of occlusion force necessary to fully seal the tube. This defect prevents the pump from complete occlusion and results in the potential for over-infusion of drug product into the patient.
Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.
What was recalled
Centurion Convenience Kits, containing BD Alaris Pump Infusion Sets. BD code 11522558 (Centurion code 110931) contained inside kits DYNJ52509A and DYNJ52510A. Product Usage: The Centurion Primary Set Pack and Primary Warmer Pack Convenience Kit helps continuously or intermittently deliver fluids, medications, blood and blood products to adult, pediatric or neonatal patients.
Get recall alerts
Free email alert whenever Medline Industries Inc has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Medline Industries Inc