LASER CDS recalled over sterility concerns
- Recall date
- April 12, 2017
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Official notice title
- Medline Industries, Inc. recalls LASER CDS
- Recall number
- Z-2980-2018
- FDA classification
- Class II
- Brand / firm
- Medline Industries, Inc.
- Sold / distributed
- CA, NM
Why it was recalled
Multiple surgical packs were assembled with a non-sterile CHG applicators labeled as "ChloraPrep" (kit component numbers: 29183, 52115, 82279, and 82281). The non-sterile components 29183, 52115, 82279, and 82281 should have been placed inside the sterile portion of the pack, where it would subsequently be exposed to the ethylene oxide sterilization process. However, the components were not placed in the sterile portion of the pack during assembly and were never exposed to the sterilization process. Components 29183, 52115, 82279, and 82281 were inadvertently released non-sterile. The remainder of the pack components were properly sterilized and only components 29183, 52115, 82279, and 82281 were affected. Components 29183, 52115, 82279, and 82281 are manufactured by Becton, Dickinson (BD) and are labeled "ChloraPrep". ChloraPrep" applicators contain 2% w/v chlorhexidine gluconate (CHG) in 70% v/v isopropyl alcohol (IPA). The applicators are used to facilitate skin preparation for various preoperative and pre-injection procedures.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
LASER CDS
Get recall alerts
Free email alert whenever Medline Industries, Inc. has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Medline Industries, Inc.