Medical device recalls Moderate risk

LASER CDS recalled over sterility concerns

Recall date
April 12, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Official notice title
Medline Industries, Inc. recalls LASER CDS
Recall number
Z-2980-2018
FDA classification
Class II
Brand / firm
Medline Industries, Inc.
Sold / distributed
CA, NM

Why it was recalled

Multiple surgical packs were assembled with a non-sterile CHG applicators labeled as "ChloraPrep" (kit component numbers: 29183, 52115, 82279, and 82281). The non-sterile components 29183, 52115, 82279, and 82281 should have been placed inside the sterile portion of the pack, where it would subsequently be exposed to the ethylene oxide sterilization process. However, the components were not placed in the sterile portion of the pack during assembly and were never exposed to the sterilization process. Components 29183, 52115, 82279, and 82281 were inadvertently released non-sterile. The remainder of the pack components were properly sterilized and only components 29183, 52115, 82279, and 82281 were affected. Components 29183, 52115, 82279, and 82281 are manufactured by Becton, Dickinson (BD) and are labeled "ChloraPrep". ChloraPrep" applicators contain 2% w/v chlorhexidine gluconate (CHG) in 70% v/v isopropyl alcohol (IPA). The applicators are used to facilitate skin preparation for various preoperative and pre-injection procedures.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

LASER CDS

Get recall alerts

Free email alert whenever Medline Industries, Inc. has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Medline Industries, Inc.