Medical device recalls Moderate risk

Medline Industries Inc recalls Medicine Cup, 1 OZ. Liquid medication dispenser to issue a measured amount of liquid medication.

Recall date
May 11, 2016
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2404-2016
FDA classification
Class II
Brand / firm
Medline Industries Inc
Sold / distributed
Continental U.S.

Why it was recalled

There is 5 mL printed next to the increment on the medicine cups instead of 10 mL. The cup has the correct mL markings for the other gradations, but sequentially, the markings are listed as 2.5 mL, 5 mL, 7.5 mL, and 5 mL.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Medicine Cup, 1 OZ. Liquid medication dispenser to issue a measured amount of liquid medication.