Medline Industries Inc recalls Medicine Cup, 1 OZ. Liquid medication dispenser to issue a measured amount of liquid medication.
- Recall date
- May 11, 2016
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2404-2016
- FDA classification
- Class II
- Brand / firm
- Medline Industries Inc
- Sold / distributed
- Continental U.S.
Why it was recalled
There is 5 mL printed next to the increment on the medicine cups instead of 10 mL. The cup has the correct mL markings for the other gradations, but sequentially, the markings are listed as 2.5 mL, 5 mL, 7.5 mL, and 5 mL.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Medicine Cup, 1 OZ. Liquid medication dispenser to issue a measured amount of liquid medication.