Medline brand medical procedure convenience kits recalled over sterility concerns
- Recall date
- September 4, 2024
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Official notice title
- MEDLINE INDUSTRIES, LP - Northfield recalls Medline brand medical procedure convenience kits, labeled as: a) OB PACK-LF, REF DYNJ0912449M; b) OB PACK, REF DYNJ1730…
- Recall number
- Z-0177-2025
- FDA classification
- Class II
- Brand / firm
- MEDLINE INDUSTRIES, LP - Northfield
- Sold / distributed
- Worldwide distribution - US Nationwide and the countries of Canada, Panama.
Why it was recalled
It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Medline brand medical procedure convenience kits, labeled as: a) OB PACK-LF, REF DYNJ0912449M; b) OB PACK, REF DYNJ17309G; c) OB PK, REF DYNJ34268D; d) OB DELIVERY PACK, REF DYNJ34274B; e) OB DELIVERY PACK, REF DYNJ34274B; f) D AND C PACK, REF DYNJ39668I; g) OB PACK, REF DYNJ42884; h) RFT D & C PACK, REF DYNJ50851; i) KIT OB, REF DYNJ906623; j) D AND C, REF DYNJ907957A; k) CHS OB PACK III, REF DYNJCH907
Get recall alerts
Free email alert whenever MEDLINE INDUSTRIES, LP - Northfield has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: MEDLINE INDUSTRIES, LP - Northfield