Medical device recalls Moderate risk

Medline Industries, LP recalls Convenience kits containing Tego Connectors Medline DIALYSIS DRSNG CHNG SKU DYNDC3167

Recall date
January 8, 2026
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1468-2026
FDA classification
Class II
Brand / firm
Medline Industries, LP
Sold / distributed
US Nationwide distribution.

Why it was recalled

Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Convenience kits containing Tego Connectors Medline DIALYSIS DRSNG CHNG SKU DYNDC3167

Get recall alerts

Free email alert whenever Medline Industries, LP has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Medline Industries, LP