Medical device recalls Moderate risk

Medline Industries, LP recalls Medline Industries IV Administration and Extension Sets: 1) DBD-IV ADMIN SET,60DRP,1NEEDLE-FREE VALV, Model Number: DYN…

Recall date
December 31, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1366-2026
FDA classification
Class II
Brand / firm
Medline Industries, LP
Sold / distributed
Worldwide distribution - US Nationwide and the country of Bahamas.

Why it was recalled

Under circumstances involving exposure to particular chemicals and mechanical forces, the male luer connectors for the specified IV Administration and Extension sets may leak, crack, and/or break during use, which may result in a delay in patient treatment, blood loss, infection, an air embolism, and/or underdosage of medication.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Medline Industries IV Administration and Extension Sets: 1) DBD-IV ADMIN SET,60DRP,1NEEDLE-FREE VALV, Model Number: DYNDTB0529; 2) IV ADMIN SET, 10DRP, 1NF, 92, Model Number: DYNDTB1029; 3) DBD-IVEXTSET2.4ML2 NEEDLE-FREEVLVE,35, Model Number: DYNDTB1516; 4) DBD-IV ADMIN SET, 15DRP, 1 INJECT PORT,, Model Number: DYNDTB1530; 5) DBD-IV ADMIN SET, 15DRP, ROLLER CLAMP, 6, Model Number: DYNDTB1539; 6) DBD-IV ADMIN SET, 15DRP, 3 NF, 109, Model Number: DYNDTB1545; 7) IV ADMIN SET, 20DRP, 3 NF, 109, Model Number: DYNDTB2045; 8) DBD-IVADMINSET60DRP2NEEDLE-FREEVLVE96, Model Number: DYNDTB0512; 9) DBD-IVADMINSET,60DRP3 NEEDLE-FREE VLVE11, Model Number: DYNDTB0555; 10) IV ADMIN SET, 10DRP, NF,INJ, 96, Model Number: DYNDTB1022; 11) DBD-IV ADMIN SET, 15DRP, 2NF, 96, Model Number: DYNDTB1512; 12) DBD-IV ADMIN SET, 15DRP, 1NF, 92, Model Number: DYNDTB1529; 13) DBD-IV ADMIN SET, 15DRP, ROLLER CLAMP, 9, Model Number: DYNDTB1537; 14) DBD-IV ADMIN SET, 15DRP, SECOND, 40, Model Number: DYNDTB1540; 15) IV ADMIN SET, 20DRP, 1NF, 92, Model Number: DYNDTB2029; 16) DBD-IV EXT SET, 6.5, 1.1ML, M LUERLOCK,, Model Number: DYNDTB5077; 17) IV ADMIN SET, 10DRP, 3 NF, 109, Model Number: DYNDTB1045; 18) DBD-IVADMISET15DRP4NEEDLEFREE VLVESTPCK1, Model Number: DYNDTB1545D; 19) IV ADMIN SET15DRPFLWCTRL1NEEDLEFREE VA, Model Number: DYNDTB2524; 20) IV ADMIN SET, 10DRP, 2NF, 96, Model Number: DYNDTB1012; 21) IV ADMIN SET, 20DRP, 2NF, 96, Model Number: DYNDTB2012; 22) IV ADMIN SET, 20DRP, NF,INJ, 96, Model Number: DYNDTB2022; 23) DBD-IV EXT SET,SMBORE EXT,8,M LUERLOCK,1, Model Number: DYNDTB5081