Medline Industries, LP recalls Medline medical procedure convenience kits labeled as: 1) THORACIC ROBOTICS, REF DYNJ908777B; 2) CARDIAC ROBOTIC WHC KI…
- Recall date
- December 16, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1085-2026
- FDA classification
- Class II
- Brand / firm
- Medline Industries, LP
- Sold / distributed
- US Nationwide distribution.
Why it was recalled
Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Medline medical procedure convenience kits labeled as: 1) THORACIC ROBOTICS, REF DYNJ908777B; 2) CARDIAC ROBOTIC WHC KIT-LF, REF DYNJ911359
Get recall alerts
Free email alert whenever Medline Industries, LP has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Medline Industries, LP