Medical device recalls Moderate risk

Medline Industries, LP recalls Medline medical procedure convenience kits labeled as: KIT ROBOTICS UROLOGY PROSTATE, REF DYKMBNDL116G

Recall date
December 16, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1090-2026
FDA classification
Class II
Brand / firm
Medline Industries, LP
Sold / distributed
US Nationwide distribution.

Why it was recalled

Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Medline medical procedure convenience kits labeled as: KIT ROBOTICS UROLOGY PROSTATE, REF DYKMBNDL116G

Get recall alerts

Free email alert whenever Medline Industries, LP has a new recall — straight from official government data. Unsubscribe anytime.