MEDTECH SAS recalls ROSA One 3.1 Brain application The device is intended for the spatial positioning and orientation of instruments holder…
- Recall date
- September 22, 2021
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0118-2022
- FDA classification
- Class I
- Brand / firm
- MEDTECH SAS
- Sold / distributed
- US Nationwide distribution and International Distribution to countries of: Australia, Taiwan, China, Thailand, Japan, India, South Korea, and the Netherlands.
Why it was recalled
The firm has become aware of a software anomaly affecting ROSA One 3.1 Brain application which led to the inaccurate placement of an electrode during surgery. The firm has received 3 global complaints related to the issue. An incorrect trajectory could result in serious injury or death if undetected during surgery.
Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.
What was recalled
ROSA One 3.1 Brain application The device is intended for the spatial positioning and orientation of instruments holders or tool guides to be used by trained surgeons to guide standard surgical instruments during brain and spine surgeries.
Get recall alerts
Free email alert whenever MEDTECH SAS has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: MEDTECH SAS