Medical device recalls Moderate risk

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalls Medtronic Power Supply 26907, Medtronic CareLink Encore(TM) 29901 Programmer

Recall date
January 17, 2019
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0845-2019
FDA classification
Class II
Brand / firm
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Sold / distributed
US and Armenia, Bulgaria, Croatia, Georgia, Hungary, India, Iran, Jordan, Kazakhstan, Latvia, Lebanon, Oman, Pakistan, Qatar, Romania, Saudi Arabia, Syrian Arab Republic

Why it was recalled

There is a potential for an electrical short circuit to develop in a subset of 479 Model 26907 power supply cable plugs that connect to the Medtronic Model 29901 CareLink Encore programmer. A failure mode of the power supply has been found where an electrical short occurs on the low voltage direct current connector and results in the connector heating up, generating smoke, melting and in some cases disconnecting from the programmer.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Medtronic Power Supply 26907, Medtronic CareLink Encore(TM) 29901 Programmer

Get recall alerts

Free email alert whenever Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) has a new recall — straight from official government data. Unsubscribe anytime.