Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalls Medtronic Power Supply 26907, Medtronic CareLink Encore(TM) 29901 Programmer
- Recall date
- January 17, 2019
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0845-2019
- FDA classification
- Class II
- Brand / firm
- Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
- Sold / distributed
- US and Armenia, Bulgaria, Croatia, Georgia, Hungary, India, Iran, Jordan, Kazakhstan, Latvia, Lebanon, Oman, Pakistan, Qatar, Romania, Saudi Arabia, Syrian Arab Republic
Why it was recalled
There is a potential for an electrical short circuit to develop in a subset of 479 Model 26907 power supply cable plugs that connect to the Medtronic Model 29901 CareLink Encore programmer. A failure mode of the power supply has been found where an electrical short occurs on the low voltage direct current connector and results in the connector heating up, generating smoke, melting and in some cases disconnecting from the programmer.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Medtronic Power Supply 26907, Medtronic CareLink Encore(TM) 29901 Programmer
Get recall alerts
Free email alert whenever Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) has a new recall — straight from official government data. Unsubscribe anytime.