Medical device recalls Moderate risk

Medtronic Inc., Cardiac Rhythm and Heart Failure recalls Medtronic, Viva Quadripolar Implantable Cardioverter Defibrillators: Product Model VIVAXTCRT-D DTBA1D1, DTBA1D4 VIVASCR…

Recall date
August 12, 2016
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2650-2016
FDA classification
Class II
Brand / firm
Medtronic Inc., Cardiac Rhythm and Heart Failure
Sold / distributed
Worldwide Distribution - US Nationwide in the states of US: AZ, CA, FL, GA, IL, KY, MD, MN, NJ, NY, NC, OH, OK, PA, PR, TX. and countries of: Australia, Finland, France, Germany, Iran, Italy, Japan, Korea, Netherlands, Puerto Rico, Saudi Arabia, Spain, and United Kingdom.

Why it was recalled

78 Viva CRT -Ds and Evera ICDs that were manufactured with a specific subset of circuit components may experience rapid battery depletion due to a low resistance path developing within the circuit component. There have been seven (7) confirmed failures (9%) through July 16, 2016. Reported complications have included shortness of breath, pocket heating, low heart rate, and early device explant.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Medtronic, Viva Quadripolar Implantable Cardioverter Defibrillators: Product Model VIVAXTCRT-D DTBA1D1, DTBA1D4 VIVASCRT-D DTBB1D1, DTBB1D4 Product Usage: The Medtronic Viva CRT -D dual chamber, implantable cardioverter defibrillator with cardiac resynchronization therapy (CRT -D) is a multiprogrammable cardiac device that monitors and regulates the patients heart rate by providing single or dual chamber, rate responsive bradycardia pacing; sequential biventricular pacing; ventricular tachyarrhythmia therapies; and atrial tachyarrhythmia therapies.

Get recall alerts

Free email alert whenever Medtronic Inc., Cardiac Rhythm and Heart Failure has a new recall — straight from official government data. Unsubscribe anytime.