Medtronic Inc. recalls Guardian Connect App CSS7200 used on the iPhone, iPad, and iPod Touch devices.
- Recall date
- July 4, 2018
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1282-2019
- FDA classification
- Class II
- Brand / firm
- Medtronic Inc.
- Sold / distributed
- Distribution was only made OUS. Countries involved are: Austria, United Arab Emirate, Australia, Belgium, Switzerland, Chile, Colombia, Czech Republic, Germany, Denmark, Spain, Finland, France, United Kingdom, Hungary, Ireland, Israel, Iceland, Italy, Republic of Korea, Luxembourg, Netherlands, Nor…
Why it was recalled
The application may be closed by the operating system without alerting the user the app is no longer running or communicating with the transmitter resulting in the user not receiving alerts that could be associated with hypoglycemic or hyperglycemic events.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Guardian Connect App CSS7200 used on the iPhone, iPad, and iPod Touch devices.