Medical device recalls Moderate risk

Medtronic Inc. recalls Medtronic Implantable Insulin Pump System, REF: MMT-2007D, Sterile EO, CE 0459 UDI: 00763000043001

Recall date
April 10, 2020
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2206-2020
FDA classification
Class II
Brand / firm
Medtronic Inc.
Sold / distributed
No US Distribution. OUS: Netherlands, France, and Sweden

Why it was recalled

Internal testing results showed endotoxin-related test result did not meet specifications for implantable insulin pump. If pumps have higher than allowed endotoxin levels, patients may have an effect of fever, hypotension, anaphylactic shock, hyperglycemia, or diabetic ketoacidosis (DKA), potentially followed by secondary multi-organ failure (primarily renal and hepatic) and/or death.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Medtronic Implantable Insulin Pump System, REF: MMT-2007D, Sterile EO, CE 0459 UDI: 00763000043001