Medical device recalls Moderate risk

Medtronic MiniMed Inc. recalls MiniMed NGP 640G 1.8ml (mmol/L), Model No. MMT-1511; NGP 640G 1.8ml (mg/dL), Model No. 1512; NGP 640G PLGM 3ml (mmol/L)…

Recall date
June 19, 2015
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2260-2015
FDA classification
Class II
Brand / firm
Medtronic MiniMed Inc.
Sold / distributed
Australia, Czech Republic, Denmark, Finland, France, Ireland, Netherlands, Norway, Sweden, United Kingdom.

Why it was recalled

Medtronic MiniMed is recalling the MiniMed 620G and 640G insulin pumps because there are certain scenarios where the set Bolus screen will not timeout, which could cause confusion by showing a bolus amount that is no longer appropriate.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

MiniMed NGP 640G 1.8ml (mmol/L), Model No. MMT-1511; NGP 640G 1.8ml (mg/dL), Model No. 1512; NGP 640G PLGM 3ml (mmol/L), Model No. MMT-1711; NGP 640G PLGM 3ml (mg/dL), Model No. MMT-1712.

Get recall alerts

Free email alert whenever Medtronic MiniMed Inc. has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Medtronic MiniMed Inc.