Medtronic Minimed recalls MiniMed 780G Insulin Pump: Models: PUMP MMT-1885 MM780G 6.5V BLE MMOL;¿ PUMP MMT-1886 MM780G 6.5W BLE MG; KIT MMT-1895W…
- Recall date
- March 15, 2021
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1771-2021
- FDA classification
- Class II
- Brand / firm
- Medtronic Minimed
- Sold / distributed
- U.S.(Clinical): AR, CA, CO, CT, FL, GA, ID, MI, MN, TN, TX, and WA O.U.S.: Austria, Bahrain, Belgium, Canary Islands, Chile, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Gibraltar, Greece, Hungary, Iceland, Ireland, Israel, Italy, Liechtenstein, Luxembourg, Malta, Martinique,…
Why it was recalled
Due to a software design issue, under certain conditions, a software fault is detected when a large bolus delivery at quick bolus speed completes. if the user is not aware of the amount of active insulin and delivers an additional bolus, there is a risk of insulin over delivery.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
MiniMed 780G Insulin Pump: Models: PUMP MMT-1885 MM780G 6.5V BLE MMOL;¿ PUMP MMT-1886 MM780G 6.5W BLE MG; KIT MMT-1895WW MM780G 6.5V BLE SF MMOL;¿ KIT MMT-1896WW MM780G 6.5W BLE SF MG;¿ PUMP MMT-1884XC 780G V6.5 CLIN US MG;¿ PUMP MMT-1885XC 780G V6.5 CLIN MM;¿ PUMP MMT-1886XC 780G V6.5 CLIN MG;¿ PUMP MMT-1885L MM780G 6.5V BLE MMOL;¿ PUMP MMT-1886L MM780G 6.5W BLE MG;¿ KIT MMT-1896ES MM780G V6.5W MG;¿ KIT MMT-1896WWA MM780G BLE MG ES;¿ KIT MMT-1895WWA MM780G BLE MMOL AR;¿ KIT MMT-1895WWA MM780G BLE MMOL CS;¿ KIT MMT-1895WWA MM780G BLE MMOL DA;¿ KIT MMT-1895WWA MM780G BLE MMOL DE;¿ KIT MMT-1895WWA MM780G BLE MMOL EN;¿ KIT MMT-1895WWA MM780G BLE MMOL FI;¿ KIT MMT-1895WWA MM780G BLE MMOL FR;¿ KIT MMT-1895WWA MM780G BLE MMOL IT;¿ KIT MMT-1895WWA MM780G BLE MMOL NL;¿ KIT MMT-1895WWA MM780G BLE MMOL NO;¿ KIT MMT-1895WWA MM780G BLE MMOL SL; KIT MMT-1895WWA MM780G BLE MMOL SV;¿ KIT MMT-1895WWA MM780G BLE MMOL HU;¿ KIT MMT-1896WWA MM780G BLE MG AR;¿ KIT MMT-1896WWA MM780G BLE MG DE;¿ KIT MMT-1896WWA MM780G BLE MG EL;¿ KIT MMT-1896WWA MM780G BLE MG HE;¿ KIT MMT-1896WWA MM780G BLE MG IT;¿ KIT MMT-1896WWA MM780G BLE MG PL;¿ KIT MMT-1896WWA MM780G BLE MG RO;¿ KIT MMT-1896WWA MM780G BLE MG TR;¿ KIT MMT-1896WWA MM780G BLE MG NL;¿ KIT MMT-1896WWA MM780G BLE MG FR;¿ KIT MMT-1896WWA MM780G BLE MG ES;
Get recall alerts
Free email alert whenever Medtronic Minimed has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Medtronic Minimed