Medtronic Navigation, Inc. recalls Medtronic Instrument Set EM ENT, Ref. #9733452 and #9733452-G02, reusable, RX. This kit contains 9733449, Straight Suct…
- Recall date
- June 29, 2017
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-3055-2017
- FDA classification
- Class II
- Brand / firm
- Medtronic Navigation, Inc.
- Sold / distributed
- Worldwide Distribution - US (nationwide) and Internationally to Canada.
Why it was recalled
The instruments are not able to pass instrument verification which is performed prior to the use of each instrument.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Medtronic Instrument Set EM ENT, Ref. #9733452 and #9733452-G02, reusable, RX. This kit contains 9733449, Straight Suction, 9733450, 70 deg Curved Suction, and 9733451, 90 deg Curved Suction. The firm name on the label is Medtronic Navigation, Inc., Louisville, CO. The suction devices are part of a family of reusable ENT instruments that are used with the StealthStationTM AxiEM surgical navigation system. The AxiEM system is an electromagnetic (EM) tracking solution for surgical tools. The system measures an induced voltage created by local magnetic fields from a set of coils in the transmit coil array (TCA). Each surgical tool has a set of receiving coils to sense the magnetic field strength. Using a mathematical algorithm, the position and orientation of the instrument can be computed and overlaid/registered to imagery on a computer monitor, resulting in the ability to track the instrument tip in relation to the patients anatomy. AxiEM ENT Suction devices are used as suction and pointing devices. The suction devices are intended to have an ENT Instrument Tracker mounted on it
Get recall alerts
Free email alert whenever Medtronic Navigation, Inc. has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Medtronic Navigation, Inc.