Medical device recalls High risk

Medtronic Navigation, Inc. recalls Medtronic Nexframe Stereotactic System and StealthStation Cranial software version 3.0 or newer with StealthStation DBS…

Recall date
August 9, 2019
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1995-2020
FDA classification
Class I
Brand / firm
Medtronic Navigation, Inc.
Sold / distributed
Worldwide distribution including US nationwide, Austria, Canada, Italy, Romania, Spain, and United Kingdom.

Why it was recalled

Entry point and lead placement inaccuracies during deep brain stimulation lead implantation procedures may occur when using a specific combination of the firm's Steriotactic System and auto-registration feature with a specific imaging system (also known as a fiducial-less procedure). Minor patient movement may not be initially detected by the user or the software during the auto-registration scan process potentially resulting in inaccuracies and risks for the patient including: inaccurate lead placement, delay of surgery, aborted surgery, or additional intervention (including revision of the lead placement and subsequent imaging).

Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.

What was recalled

Medtronic Nexframe Stereotactic System and StealthStation Cranial software version 3.0 or newer with StealthStation DBS License or StealthStation S8 Software with Stealth DBS License and O-arm Imaging System utilizing the auto-registration (fiducial- less) workflow used in combination during a DBS (deep brain stimulation) procedure. Neurological stereotaxic Instrument.