Medical device recalls Moderate risk

Medtronic Navigation, Inc. recalls Spine Referencing Instrumentation, Spinous Process Clamp, Short When used with a Medtronic StealthStation Navigation Sy…

Recall date
June 29, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-3140-2017
FDA classification
Class II
Brand / firm
Medtronic Navigation, Inc.
Sold / distributed
Worldwide Distribution - US including AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, AD, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV, & WY. Foreign: Albania, Australia, Bangladesh, Belgium, Bulgaria, Canada, Croatia, Cz…

Why it was recalled

Potential for Spinal Clamps to be damaged when opened beyond intended limits, either no opening once attached to a patient's spinous process or leaving a component behind in the patient's body following breakage.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Spine Referencing Instrumentation, Spinous Process Clamp, Short When used with a Medtronic StealthStation Navigation System, the Spine Referencing fixation devices are intended to provide rigid attachment between patient and patient reference frame for the duration of the surgery. The devices are intended to be reusable. The spine referencing clamps provide bony fixation through two styles of clamps; a single spinous process clamp and a double spinous process clamp.