Medtronic Neuromodulation recalls DBS Pocket Adaptor, Model 64001 and Model 64002. The Medtronic Models 64001 (1x4) and 64002 (2x4) Pocket Adaptors can b…
- Recall date
- June 23, 2016
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2645-2016
- FDA classification
- Class II
- Brand / firm
- Medtronic Neuromodulation
- Sold / distributed
- Worldwide Distribution.
Why it was recalled
Due to the potential for pocket adaptor conductor wire fractures involving DBS Pocket Adaptor Model 64001 and Model 64002, a safety notification was sent to healthcare professionals to reinforce device specific labeling for the handling of pocket adaptors and system integrity checking during implant procedures.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
DBS Pocket Adaptor, Model 64001 and Model 64002. The Medtronic Models 64001 (1x4) and 64002 (2x4) Pocket Adaptors can be used as a part of a neurostimulation system for deep brain stimulation. The pocket adaptor is intended to be implanted with the new replacement neurostimulator in the same pocket used for the explanted neurostimulator. Implanting in the same neurostimulator pocket allows for a single-incision procedure.
Get recall alerts
Free email alert whenever Medtronic Neuromodulation has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Medtronic Neuromodulation