Medical device recalls Moderate risk

Medtronic Sofamor Danek USA Inc recalls Medtronic ZEVO(TM) Anterior Cervical Plate System SLOT SCREW, REF 7710015, Size 4.5mm x 15mm Product Appliance, fixatio…

Recall date
March 23, 2018
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1750-2018
FDA classification
Class II
Brand / firm
Medtronic Sofamor Danek USA Inc
Sold / distributed
Worldwide Distribution - US Nationwide in the states of GA, CT, LA, MD, WY, IN, MD, FL, and France

Why it was recalled

Slot screw had an overall length of 15mm, but per the specification, the screw should have an overall length of approximately 17mm.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Medtronic ZEVO(TM) Anterior Cervical Plate System SLOT SCREW, REF 7710015, Size 4.5mm x 15mm Product Appliance, fixation, spinal intervertebral body

Get recall alerts

Free email alert whenever Medtronic Sofamor Danek USA Inc has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Medtronic Sofamor Danek USA Inc