Merge Healthcare, Inc. recalls The Schiller PB 1000, also known as Merge Hemo Patient Data Module (PDM), a major component of the Merge Hemo software.
- Recall date
- March 10, 2016
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2707-2017
- FDA classification
- Class II
- Brand / firm
- Merge Healthcare, Inc.
- Sold / distributed
- Distribution US nationwide.
Why it was recalled
If the PDM is utilized to output a pressure for computing fractional flow reserve (FFR), the analog output may not reflect the true representation of the selected pressure waveform being provided by the PDM to the Link Assembly which provides the analog output.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
The Schiller PB 1000, also known as Merge Hemo Patient Data Module (PDM), a major component of the Merge Hemo software.
Get recall alerts
Free email alert whenever Merge Healthcare, Inc. has a new recall — straight from official government data. Unsubscribe anytime.