Medical device recalls High risk

Merit Medical Systems, Inc. recalls 7F Prelude¿ Short Sheath Introducer. Catalog Numbers: K15-00070, K15-00170, PSS-7F-4-035MT, PSS-7F-4MT. Produc Usage: T…

Recall date
February 16, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1497-2017
FDA classification
Class I
Brand / firm
Merit Medical Systems, Inc.
Sold / distributed
Nationwide and other foreign consignees. No Canadian or VA/govt/military consignees.

Why it was recalled

Merit Medical Systems, Inc. announces a voluntary field action for sheaths included with the 7F Prelude¿ Short Sheath Introducers because the sheath marker tips may detach during use.

Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.

What was recalled

7F Prelude¿ Short Sheath Introducer. Catalog Numbers: K15-00070, K15-00170, PSS-7F-4-035MT, PSS-7F-4MT. Produc Usage: The Merit Prelude¿ Short Sheath Introducer is intended to provide access and facilitate the percutaneous introduction of various devices into veins and/or arteries while maintaining hemostasis for a variety of diagnostic and therapeutic procedures. The device can also provide access to a native or synthetic graft used for hemodialysis. The side port of the sheath allows adequate flow to perform temporary hemodialysis. The device is not indicated for long term vascular or hemodialysis access.

Get recall alerts

Free email alert whenever Merit Medical Systems, Inc. has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Merit Medical Systems, Inc.