Merit Medical Systems, Inc. recalls PreludeSYNC DISTAL Radial Compression Devices, SDRB-REG-LT, SDRB-REG-RT
- Recall date
- March 11, 2020
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1796-2020
- FDA classification
- Class II
- Brand / firm
- Merit Medical Systems, Inc.
- Sold / distributed
- U.S.: NV, IL, VA, PA, CA, AR, MA, GA, OH, NJ, TX, FL, NC, SC, MI, NH, CO, AL, MN, KS, NM, AZ, ID, MD, NY, OR, DC, DE, MO, IN, WI, IA, TN. O.U.S.: Switzerland, Hong Kong, Spain, Portugal, Canada, Singapore, United Arab Emirates, Mexico, United Kingdom, France, Japan, Germany, Italy
Why it was recalled
During the sterilization of radial compression devices, some of the inflatable balloons may break or burst, which may result in air leakage and the inability to maintain pressure/compression. This may result in a lack of hemostasis resulting in bleeding or a delay in treatment; the defect is likely to be visually identified by the clinician before use or identified during inflation of the balloon.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
PreludeSYNC DISTAL Radial Compression Devices, SDRB-REG-LT, SDRB-REG-RT