Micro Therapeutics, Inc. recalls Medtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-021-350-16, PED3-021-250-14, PED3-021-…
- Recall date
- January 30, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1246-2025
- FDA classification
- Class I
- Brand / firm
- Micro Therapeutics, Inc.
- Sold / distributed
- Worldwide distribution: US (nationwide): PR, OK, CA, MD, FL, NY, WA, MI, AL, NV, MN, NJ, IL, OH, AK, CT, TN, MO, TX, ID, AZ, WI, PA, OR, SD, VA, CO, GA, KY, IN, AR, MA, LA, NE, MT, ME, SC, DC, NC, ND, NM, IA, UT, KS, WV, NH, HI, RI. And OUS (foreign) countries of: Algeria, Argentina, Armenia, Austr…
Why it was recalled
Use of embolization device with diameters greater-than-or-equal-to 4mm may lead to incomplete wall apposition and/or braid deformation (fish-mouthing/braid narrowing/braid collapse) noted during procedure and post-procedure, typically noted at 6-12 month imaging follow-up, potentially can lead to thrombosis and/or serious adverse events. Higher risk in females less-than-or-equal-to 45 years of age
Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.
What was recalled
Medtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-021-350-16, PED3-021-250-14, PED3-021-350-20, PED3-021-250-20, PED3-021-300-16, PED3-021-300-20, PED3-021-275-12, PED3-021-275-16, PED3-021-325-12, PED3-021-325-14, PED3-021-325-16, PED3-021-325-20, PED3-021-250-12, PED3-021-300-14, PED3-021-250-16, PED3-021-300-12, PED3-021-275-14, PED3-021-350-14, PED3-021-350-25, PED3-021-350-12, PED3-021-250-10, PED3-021-275-20, PED3-021-350-18, PED3-021-325-18, PED3-021-275-18, PED3-021-250-18, PED3-021-300-18; and Instructions for Use, Part: M056989CDOC2