Medical device recalls Moderate risk

Microbiologics Inc recalls QC Sets and Panels, KWIK-STIK AST-GP (6 Strains) QC Set. KWIK-STIK kit that contains Catalog Number 0959P. In Vitro Dia…

Recall date
February 18, 2020
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0588-2022
FDA classification
Class II
Brand / firm
Microbiologics Inc
Sold / distributed
Worldwide Distribution. US nationwide including Puerto Rico and Guam, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia/Hrvatska, Curacao, Denmark, Ecuador, Egypt, El Salvador, Finland, France, Germany, Guatemala, Hong Kong, India, Ireland, Israel, Italy, Japa…

Why it was recalled

Tests returning incorrect susceptible results. High Level Gentamicin Synergy value was out of specification (sensitive) when analyzed on the Vitek 2 AST-GP67 card, and these lots failed the disk diffusion method used for product release.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

QC Sets and Panels, KWIK-STIK AST-GP (6 Strains) QC Set. KWIK-STIK kit that contains Catalog Number 0959P. In Vitro Diagnostic Control microorganisms

Get recall alerts

Free email alert whenever Microbiologics Inc has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Microbiologics Inc